Helix Pcd Device ( Helix Pcd Pipe )
Helix Pcd Device is using with helix pcd test strips which sterilization assusrance at higher level, daily test for pre-vacuum hospital sterilization and for benchtop steam sterilizers for steam penetration and inert gases.
How Helix Pcd Device to use ?
The test should be performed in a preheated chamber at the start of the day with only the Helix device present.
1) Pre-heat the autoclave, up to working temperature.
Note: Performing the Helix test on a cold autoclave can lead to incorrect results
2) Open the Helix Device, then take one strip from the book of 250, fold in half (test squares to the outside and adhesive backing to the inside) and insert the test strip into the slot in the device.
3) Close the Helix Device and place the Helix Device in the front of autoclave near the door.
4) Close the autoclave and process a sterilisation cycle.
5) The Helix Strip will change from a Pink to BLACK. If the whole strip is BLACK you have a successful test.
6) Date the Strip, remove adhesive backing and stick into your Book.
Which sterilization process is Helix Pcd Test and Device designed for ?
Helix Pcd Test and Device are designed to be used in Steam sterilizers.
Which sterilization cycle are Helix Pcd Test and Device working with ?
Helix Pcd Test and Device are designed operate at steam cycle 134 C – 3,5 minutes or 121 C – 15 minutes conditions.
What is the standart of ISO for Helix Pcd Test and Device ?
Helix Pcd Test and Device according to ISO 11140-1, ISO 11140-3 and ISO 11140-4 standart. CE marking is not applicable for indicator products, because this product is not considered a medical device under the clause of the EC-directive.
Helix Plastic Pipe
Helix Steel Pipe